OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE PRELOADED SPHINCTEROTOME V (DISTAL WIREGUIDED) Back to Search Results
Model Number KD-VC631Q-07301A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, because the subject device was discarded by the user.Therefore, the exact cause of the reported event could not be conclusively determined.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event or Problem Description
During an endoscopic sphincterotomy, three devices were used.The cutting wire of the first device broke.The user exchanged it for the second device and continued the procedure.The cutting wire of the second device broke.The user exchanged it for the third device and completed the procedure.There was no patient injury reported.This is the report regarding the breakage of the cutting wire of the second device.
 
Additional Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, because the subject device was discarded by the user.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that a part of the cutting wire became extremely hot and broke due to any of the following possible causes.The contact length of the cutting wire and the tissue or the distal part of the endoscope was too short while the subject device was activated, and spark occurred.The coating was damaged due to contacting with the metal part of the forceps elevator of the endoscope.The exposed cutting wire from the damaged coating contacted to the metal part of the forceps elevator while the device was activated, and it caused spark.The above device handling has warned in the instruction manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V (DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10500652
Report Number8010047-2020-06268
Device Sequence Number3222720
Product Code KNS
UDI-Device Identifier04953170371516
UDI-Public04953170371516
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC631Q-07301A
Device Lot Number05V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/25/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date09/08/2020
Supplement Report FDA Received Date10/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-