• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901A
Device Problems No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Computer Operating System Problem (2898)
Patient Problem No Patient Involvement (2645)
Event Date 08/20/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the screen of the programmer was not responding to touch or stylus.It was noted that the stylus would not select in the upper right part of the screen and could not expand the electrocardiogram (ecg) dialog box - the arrow jumped away.There was no visible damage found on the display.The display was replaced.Analysis was however unable to confirm or reproduce the customer comment that the programmer would not work when electrocardiogram- ekg cables were plugged into it with no image appearing on the screen.There were no cables returned.All found defective parts were replaced and all other identified issues were resolved.Reconfigured hard drive, reloaded and updated software and the device then passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the screen of the programmer was not responding to touch or stylus.The stylus was not working and the programmer would not work when electrocardiogram (ekg) cables were plugged into it with no image appearing on the screen.The programmer was received into service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10501090
MDR Text Key205916466
Report Number3004593495-2020-01017
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169763548
UDI-Public00643169763548
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-