Concomitant medical products: other relevant device(s) are: product id: ab46, serial/lot #: (b)(4), ubd: 23-mar-2022, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: ab46, serial/lot #: (b)(6), ubd: 23-mar-2022, udi#: (b)(4).Additional information received: iit was reported that 3 reliant devices were used during the procedure product analysis, three devices were received for analysis: device 1: a leak was from the mid-section of the balloon on attempted inflation.A complete radial tear was observed to the balloon.The balloon material was jagged and uneven along the tear indicating the material had been damaged prior to separation.The reported balloon burst was confirmed through analysis.Device 2: a leak was from the mid-section of the balloon on attempted inflation.A pin hole was observed on the mid-section of the balloon.Device 3: a leak was from the mid-section of the balloon on attempted inflation.A pin hole was observed on the mid-section of the balloon.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|