• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2020
Event Type  malfunction  
Manufacturer Narrative
Ge healthcares investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Unique identifier: (b)(4).
 
Event Description
The hospital reported the display arm "came down".There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The display arm will be replaced to resolve the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
MDR Report Key10501494
MDR Text Key205945791
Report Number9710602-2020-00269
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-