| Model Number |
0502477031 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a poor image.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not received for investigation at stryker endoscopy because it was repaired locally in poland.The technical service report is attached, and indicates: broken inner lens, hazy.Probable root cause: poor mate between cannula and endoscope; poor fit between endoscope needle and cannula; use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a poor image.
|
| |
|