Information was received from healthcare provider via field representative regarding instrument involved in oblique lateral interbody fusion (olif) procedure at l2 and l3.It was reported that during procedure, threaded tip broke off during placement of implant.Reported instrument broke and they were used quite frequently.They completed the procedure with back up sets that was available.Reported, they were able to remove the tip out of the instrument with needle nose pliers.So, there were no fragments left in the patient.There was delay in procedure for about 30 minutes total.There were no patient symptoms or complications reported as a result of the event.Patient is alive and no injury.
|
|
Additional information: d9, h3, h6.H3: device evaluation summary: visual and optical inspection revealed the instrument is damaged around the inserter attachment area.The edges have been rolled over and deformed along with multiple witness marks consistent with impacts from a hammer.This type of damage is consistent with material overload from multiple impacts.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|