• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3889-28
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a consumer via a healthcare professional regarding a patient who was implanted with a neurostimulator (ins) for gastrointestinal/pelvic floor and fecal incontinence.It was reported that the insulation detached from the proximal end of the wire.The leads insulation detached from the proximal end of the electrodes.The wire was pulled and replaced.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received from a healthcare professional (hcp) via a manufacturer representative (rep).It was reported that the cause of how the insulated coating came apart was unknown, but they did know it happened when the boot was pressed over the connection piece.
 
Additional Manufacturer Narrative
H3: the returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified the outer insulation of the lead was separated at the butt joint near the proximal end of the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10502789
Report Number2649622-2020-17180
Device Sequence Number3118648
Product Code EZW
UDI-Device Identifier00643169864436
UDI-Public00643169864436
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P080025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2024
Device Model Number3889-28
Device Catalogue Number3889-28
Device Lot NumberVA26JK6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2020
Supplement Date Received by Manufacturer09/08/2020
10/13/2020
Initial Report FDA Received Date09/08/2020
Supplement Report FDA Received Date09/14/2020
11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
-
-