SMITH & NEPHEW, INC. ANTHOLOGY STD BROACH SIZE 1; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71365601 |
Device Problem
Inadequacy of Device Shape and/or Size (1583)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/11/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during procedure, there was an issue with the implant sitting high after broaching.Procedure was concluded with the same devices, without a delay or and injury reported.
|
|
Manufacturer Narrative
|
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure mode but could confirm that the device is worn and shows signs of wear and tear.A functional evaluation of the returned product confirmed the stated failure mode.The device was paired with the mating part but could not function, rendering the device inoperable.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation close.`.
|
|
Search Alerts/Recalls
|
|
|