| Model Number |
24686 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation by mfr.: gladiator device - 7x40x75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.Blood was identified within the balloon which is evidence of a device leak.A microscopic examination identified a balloon longitudinal tear beginning at the proximal bond and extending approximately 20mm distally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.No issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 14aug2020.It was reported that crossing difficulty was encountered.The target lesion was located in the radial artery.A 7.0 x40, 75cm gladiator balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed longitudinal balloon tear.
|
| |
|
Search Alerts/Recalls
|