| Model Number |
2779 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that guide wire coating issue occurred.A 185cm str pt guidewire was selected for use.However, during procedure, the coating was peeled off, and the 2.00mm x 8mm emerge balloon catheter lumen was clogged up in that coating.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned.Visual inspection found the distal tip of the guidewire was bent.Microscopic inspection found the polymer jacket coating of the guidewire had a detached section; this damage was located approximately at 69.5 inches from the proximal end.The polymer jacket that remained attached to the distal tip was found wrinkled.
|
| |
|
Event or Problem Description
|
|
It was reported that guide wire coating issue occurred.A 185cm str pt guidewire was selected for use.However, during procedure, the coating was peeled off, and the 2.00mm x 8mm emerge balloon catheter lumen was clogged up in that coating.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|