| Model Number |
3525-0440S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348); Implant Pain (4561)
|
| Date of Event |
06/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
|
| |
|
Event or Problem Description
|
|
As reported in medwatch (b)(4): "md's description of the incident: [left] gamma3 nail that broke through proximal screw aperture at 6 weeks.Nail was noted to be notched most likely most likely at time of insertion.Nail was found to be broken in x-ray 1.5 months later." patient was revised to another nail.
|
| |
|
Event or Problem Description
|
|
As reported in medwatch (b)(4): "md's description of the incident: [left] gamma3 nail that broke through proximal screw aperture at 6 weeks.Nail was noted to be notched most likely most likely at time of insertion.Nail was.Found to be broken in x-ray 1.5 months later." patient was revised to another nail.Additionally reported: "patient was walking on memorial day, when he felt a sudden increase in pain in his left hip.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed with the help of pictures provided.A device inspection was not possible since the affected device was not returned.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.The quality certificate of the heat number was reviewed.It corresponds to the material defined on the drawing and to the internal material specification.A review of the labeling did not indicate any abnormalities.The instructions for use states: ¿avoid surface damage of implants.Discard all damaged or mishandled implants.Contouring or bending of an implant should be avoided where possible, because it may reduce its fatigue strength and can cause failure under load.If contouring is necessary, allowed by design or prescribed by stryker, the physician should avoid sharp bending, reverse bending or bending the device at a screw hole.Such action must be performed with stryker instruments and in accordance with the specified procedures (see operative technique manual).¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, one of the possible causes of the breakage could be user related i.E.Handling failure (damage of implants during implantation either during drilling or lag screw insertion).If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|