The reported event could be confirmed with the help of pictures provided.A device inspection was not possible since the affected device was not returned.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.The quality certificate of the heat number was reviewed.It corresponds to the material defined on the drawing and to the internal material specification.A review of the labeling did not indicate any abnormalities.The instructions for use states: ¿avoid surface damage of implants.Discard all damaged or mishandled implants.Contouring or bending of an implant should be avoided where possible, because it may reduce its fatigue strength and can cause failure under load.If contouring is necessary, allowed by design or prescribed by stryker, the physician should avoid sharp bending, reverse bending or bending the device at a screw hole.Such action must be performed with stryker instruments and in accordance with the specified procedures (see operative technique manual).¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, one of the possible causes of the breakage could be user related i.E.Handling failure (damage of implants during implantation either during drilling or lag screw insertion).If device is returned or any further information is provided, the investigation report will be reassessed.
|