• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PROFORE 18-25CM ANKLE CIRC. CASE 8; BANDAGE, ELASTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PROFORE 18-25CM ANKLE CIRC. CASE 8; BANDAGE, ELASTIC Back to Search Results
Catalog Number 66000016
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/31/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during treatment the profore 18-25cm ankle circ.Case 8 light conformable bandage was no rolled neatly.The treatment finished with a smith and nephew backup device.No surgical delay.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirmed the rolls where not wound neatly, establishing a relationship between the device and the reported event.The root cause identified as a missed quality check.A review of the manufacturing records found that there was no evidence that the product didn't meet specifications at the time of manufacture.A complaint history review found further instances of the reported events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFORE 18-25CM ANKLE CIRC. CASE 8
Common Device Name
BANDAGE, ELASTIC
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10505105
Report Number8043484-2020-02774
Device Sequence Number16461327
Product Code FQM
UDI-Device Identifier05000223415129
UDI-Public5000223415129
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Catalogue Number66000016
Device Lot Number966503
Initial Date Received by Manufacturer 08/13/2020
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date09/08/2020
Supplement Report FDA Received Date09/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-