| Model Number |
6248V-90 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implant procedure, the catheter broke in half during the slitting process causing the left ventricular (lv) lead to dislodge.It was noted upon dislodging, the lead grabbed the loose end of the catheter and brought it out of the body.The lv lead was successfully placed and remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the delivery catheter was returned and analyzed.The analysis indicated that the mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.There was an issue with the catheter shaft.The mechanical operation of the catheter shaft was kinked/buckled.The catheter did not peel along the score lines.Visual analysis of the lead indicated damage during use.The analyst noted the delivery catheter was returned in two pieces without the dilator.Slitting did not start on the score line.The shaft of the delivery catheter was spiral slit.The shaft of the delivery catheter was kink/buckled at 23 cm.The shaft of the delivery catheter was torn at 54.5 cm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|