• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 115-203
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii.The device has been explanted.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii.The device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the photos identified white particle materials on the shell.Device analysis performed through photographs, due to the impossibility to perform microscopic analysis as it was not possible to determine the most likely failure mode.Additional, changed, and/or corrected data: d.6b., h.6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key10506363
Report Number9617229-2020-14684
Device Sequence Number3344521
Product Code FTR
UDI-Device Identifier10888628002333
UDI-Public10888628002333
Combination Product (Y/N)N
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/10/2022
Device Model Number115-203
Device Catalogue Number115-203
Device Lot Number3022238
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer12/09/2020
Initial Report FDA Received Date09/09/2020
Supplement Report FDA Received Date01/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
-
-