• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM,,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM,, Back to Search Results
Model Number 20400109
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Fluid Discharge (2686)
Event Date 08/10/2020
Event Type  Injury  
Manufacturer Narrative
This medwatch is not to report a device malfunction, but to report an adverse patient effect post procedure.There was no report of a device malfunction or patient complication during the procedure.As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint could not be confirmed given the nature of the patient serious adverse event.There was no report of nanoknife probe or hardware malfunction during the procedure.Without receiving product to evaluate, we cannot determine if any non-conformance was present in the probe; root cause cannot be determined.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the instructions for use, which is supplied to the end user with this catalog number, contains the following statement: "warnings: do not use a device with damaged insulation.Do not attach anything to the device unless it is supplied by angiodynamics and indicated for use with this device.Attachments may damage the insulation and contribute to patient injury.Inspect all of the devices and packaging for damage prior to use.Do not use any devices that are damaged or if the sterile barrier is breached." indications for use the nanoknife system with six outputs is indicated for the surgical ablation of soft tissue.Potential adverse effects damage to critical anatomical structure (nerve, vessel, duct).A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event Description
This medwatch is not to report a device malfunction, but to report an adverse patient effect.An angiodynamics territory manager reported an issue that occurred due to a nanoknife treatment.Post procedure, the patient started having nausea/vomiting & increasing wbc count so a ct scan was performed.The scan indicated the patient was experiencing a pancreatic duct leak after treatment.A second drain was placed to drain the pool of pancreatic fluid.It was reported that is patient is okay.The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOKNIFE SYSTEM
Type of Device
LOW ENERGY DIRECT CURRENT ABLATION SYSTEM,,
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key10507152
MDR Text Key206124063
Report Number1319211-2020-00141
Device Sequence Number1
Product Code OAB
UDI-Device IdentifierH787204001090
UDI-PublicH787204001090
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number20400109
Device Catalogue Number20400109
Device Lot NumberUNKOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-