• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012625-39
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned to abbott vascular for analysis.Return of the stent delivery system (sds), inflation device and accessory devices may have further aided the analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine conclusive cause for the reported stent deployment issue since the device was not returned.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that this was a peripheral procedure to treat the heavily calcified iliac artery.The omnilink elite (ole) stent delivery system (sds) was advanced without resistance to the iliac, but the sds did not inflate.After removal, the stent was intact on the balloon sds.Another ole sds was used to complete the procedure without further incident.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10508655
Report Number2024168-2020-07489
Device Sequence Number1217539
Product Code NIO
UDI-Device Identifier08717648179341
UDI-Public08717648179341
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number1012625-39
Device Catalogue Number1012625-39
Device Lot Number9042341
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2020
Initial Report FDA Received Date09/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-