ENCORE MEDICAL L.P. BONE CEMENT; COBALT HV BONE CEMENT 40GM Back to Search Results
Model Number 600-15-000
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Date of Event 06/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Reported incident - a patient had an initial left unilateral total knee arthroplasty.Subsequently, the patient developed a superficial wound infection of the left knee.
 
Additional Manufacturer Narrative
The reason for this incident was reported as infection.The initial surgery and the time when the patient was placed on 7 days of oral keflex (antibiotic) and aquafor on wound, occurred 20 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was reported as an infection.There were no other findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE CEMENT
Common Device Name
COBALT HV BONE CEMENT 40GM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10510103
Report Number1644408-2020-00732
Device Sequence Number3134797
Product Code LOD
UDI-Device Identifier00190446138006
UDI-Public(01)00190446138006
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051496
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2020
Device Model Number600-15-000
Device Catalogue Number600-15-000
Device Lot Number844D1D0020
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/10/2020
Supplement Date Received by Manufacturer08/10/2020
Initial Report FDA Received Date09/09/2020
Supplement Report FDA Received Date02/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient Weight127
-
-