It was reported that, the "werewolf controller" would not turn on.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H10: h2, h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Visual inspection observed a damaged lid and bezel and corrosion on power supply fans.Functional evaluation revealed the unit does not turn on, the power supply led's do illuminate when plugged in, and only one power supply fan spins when plugged in.The unit was opened and found no issues inside.The complaint was confirmed and the root cause is associated with a component failure.Factors, unrelated to the manufacturing and design of the device, which could have contributed to the identified malfunction, include a defective power supply or power entry module or a blown fuse.There are no indications to suggest that the device/product did not meet specifications upon release into distribution.
|