Model Number 71692852
Medical Device Problem Codes Failure to Power Up (1476); Application Program Problem (2880)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trauma interface won't turn on.It is unknown whether the incident happened or not in a surgical environment or if there was any patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Section: h3, h6: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.There are not any signs of physical damage to the device.A functional evaluation of the returned device could not confirm the stated failure mode.The device functioned as intended.The device would turn on as well as function along with the mating components.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.H6: health effect and medical device problem code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMA INTERFACE - 2ND GEN
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key10511791
Report Number1020279-2020-04493
Device Sequence Number3134688
Product Code OLO
UDI-Device Identifier00885556584644
UDI-Public00885556584644
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170977
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71692852
Device Catalogue Number71692852
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2020
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date09/09/2020
Supplement Report FDA Received Date11/15/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1