Model Number V60
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 09sep2020.
 
Event Description
The customer contacted technical support (ts) reporting that the device is failing electrical safety testing at the proximal pressure port.The customer reported there was no patient involvement at the time the issue was discovered.The device was evaluated by the customer with assistance from the manufacturer's technical support (ts).Ts advised the customer to check connection at the grounding terminal block for the proximal port grounding wire.If the connection is bad, it is recommended to clean contacts and retest.The customer cleaned the ground terminal connected and electrical safety now passes.However, the customer reports that it is now failing high pressure leak test.Ts provided parts id for the white nylon fitting on the outlet port.Ts also recommended replacement of the blower if the issue persists.The customer stated replacement of the blower, boots, and gas delivery system (gds) resolved the issue.The customer found issue was with the gas outlet port fitting.The customer stated the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key10512053
MDR Text Key207176301
Report Number2031642-2020-03197
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/19/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-