RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 09sep2020.
|
|
Event Description
|
The customer contacted technical support (ts) reporting that the device is failing electrical safety testing at the proximal pressure port.The customer reported there was no patient involvement at the time the issue was discovered.The device was evaluated by the customer with assistance from the manufacturer's technical support (ts).Ts advised the customer to check connection at the grounding terminal block for the proximal port grounding wire.If the connection is bad, it is recommended to clean contacts and retest.The customer cleaned the ground terminal connected and electrical safety now passes.However, the customer reports that it is now failing high pressure leak test.Ts provided parts id for the white nylon fitting on the outlet port.Ts also recommended replacement of the blower if the issue persists.The customer stated replacement of the blower, boots, and gas delivery system (gds) resolved the issue.The customer found issue was with the gas outlet port fitting.The customer stated the issue was resolved.
|
|