| Catalog Number |
58-4200J-BD |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This product was being used for a trial and was from our library inventory.The product has not been returned yet for evaluation.There has been a total of (b)(4) sold of this product code with 3 complaints recorded for similar occurrences.In the 2 previous complaints, there were significant signs that the screw had been tampered with.A follow up report will be submitted one the part has been received and evaluated.
|
| |
|
Event or Problem Description
|
|
During the use of the kerrison, the screw fell out.There was a delay in the procedure to search for the screw and retrieve it.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon evalauting the returned device, it was noted that the screw was not returned with the kerrison.Without the screw it is difficult to tell if there was any damage to the screw as a result of screw removal at some point.However, there were weld marks where the back of the screw would be that indicates that the screws were properly welded into place.As a result of the evaluation, a true root cause could not be determined.This can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or need for corrective actions, a follow up report will be submitted.
|
| |
|