• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. XR ART ISRT XLPE MED SZ1-2 LT 8MM -2DEG; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. XR ART ISRT XLPE MED SZ1-2 LT 8MM -2DEG; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER Back to Search Results
Model Number 74023956
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Discomfort (2330); Injury (2348)
Date of Event 08/11/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was perform due to a feeling wrongness.S+n devices were not defective.It is unknown how is the status of the patient.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms per complaint details, ¿this case is not made by s+n defect¿ a revision was performed ¿due to a feeling of wrongness¿.Reportedly, ¿the doctor says s+n devices were not defective¿.Per correspondence, the requested medical documentation and/or explants will not be available for evaluation.S+n has not received the device and/or adequate documentation to fully evaluate the root cause of the reported revision.The patient impact beyond the reported event could not be determined.No further medical assessment is warranted at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XR ART ISRT XLPE MED SZ1-2 LT 8MM -2DEG
Common Device Name
PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10513863
Report Number1020279-2020-04501
Device Sequence Number3656116
Product Code JWH
UDI-Device Identifier00885556615942
UDI-Public00885556615942
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74023956
Device Catalogue Number74023956
Device Lot Number17AM12315
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/12/2020
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date09/10/2020
Supplement Report FDA Received Date06/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-