| Model Number |
23375 |
| Medical Device Problem Codes |
Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
08/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first day of the month of the bsc aware date.Event date not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and tip detachment occurred.The target lesion was located in the cephalic arch.A 12-4/5.8/75 xxl vascular balloon catheter was advanced for dilatation.The balloon was inflated, deflated, and was pulled down to the mid fistula.However, upon trying to inflate the balloon again, the pressure would not increase and upon pulling back the syringe, blood was noticed which indicated that the balloon had ruptured.The device was pulled out from the sheath and entire device aside from the tip was not visible.The tip lodged in an aneurysm that was successfully retrieved by making an incision and pulling it out using a curved forceps.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event - used the first day of the month of the bsc aware date.Event date not provided.Device evaluated by mfr.: the complaint device was returned for analysis.A visual examination identified a complete circumferential tear of the balloon material located approximately 10mm distal of the proximal sleeve of the balloon.The distal section of balloon material had been completely detached from the device at the distal balloon bond.This type of damage is consistent with excessive tensile force being applied to the device.The rate of burst pressure of this device is 8atm.A visual and tactile examination found the shaft to have been severely stretched/damaged between the proximal and distal balloon bonds.This type of damage is consistent with excessive tensile force being applied to the device.A visual and microscopic examination markerbands identified no issues with the markerbands.A visual and microscopic examination found no damage or issues with the tip of the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and tip detachment occurred.The target lesion was located in the cephalic arch.A 12-4/5.8/75 xxl vascular balloon catheter was advanced for dilatation.The balloon was inflated, deflated, and was pulled down to the mid fistula.However, upon trying to inflate the balloon again, the pressure would not increase and upon pulling back the syringe, blood was noticed which indicated that the balloon had ruptured.The device was pulled out from the sheath and entire device aside from the tip was not visible.The tip lodged in an aneurysm that was successfully retrieved by making an incision and pulling it out using a curved forceps.No further patient complications were reported.
|
| |
|