Catalog Number 0117180
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Defective Device (2588); Material Deformation (2976); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Hernia (2240); Injury (2348)
Date of Event 04/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol perfix light plug on (b)(6) 2016.As reported, the patient is making a claim for an adverse patient outcome against the perfix light plug.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive, no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-mar-2013) is considered to be a best estimate.Updated fields: a2, b3, b4, b5, b7, d4 (product catalog no., corporate lot no, expiry date, udi no.), d6b (date of explant), g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol perfix light plug on (b)(6) 2016.As reported, the patient is making a claim for an adverse patient outcome against the perfix light plug.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2016 ¿ patient was diagnosed with right inguinal hernia thereby underwent open repair with implant of perfix light plug.Per operative notes, ¿ direct hernia was identified and dissected.The extra large perfix light plug was placed over the defect and sutured.¿ (b)(6) 2019 - patient was diagnosed with recurrent right inguinal hernia and left inguinal hernia thereby underwent laparoscopic repair with removal of previously placed mesh and implant of two synthetic mesh.Per operative notes, ¿ large hernias were noted bilaterally.On right recurrent side, a balled mass of prosthetic mesh was identified.The mesh was seen protruding into abdominal cavity from hernia defect.The mesh was excised and defects on the left and right side was repaired with two synthetic meshes and tacked.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFIX PLUG LIGHT
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key10515033
Report Number1213643-2020-08156
Device Sequence Number17048828
Product Code FTL
UDI-Device Identifier00801741031014
UDI-Public(01)00801741031014(17)180428(10)HUXC1545
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K092032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2018
Device Catalogue Number0117180
Device Lot NumberHUXC1545
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date09/10/2020
Supplement Report FDA Received Date08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexMale
-
-