| Catalog Number |
2C8921 |
| Medical Device Problem Codes |
Material Deformation (2976); No Flow (2991)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The events occurred during the week of (b)(6) 2020.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified quantity of clearlink system 3-port manifold sets would not flow.The event was further described as tubing would kink easily and would have to be taped in several places to keep it from kinking; the primary area of kinked tubing was on each end of the manifold.The kinked tubing also prevented needed medications of saline flush from fully being administered.This was identified during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which did not identify any abnormalities that could have contributed to the reported condition.The reported condition is not clearly observed via the provided photograph and due to the nature of the returned sample no additional testing could be performed.Therefore, the reported event could not be objectively verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|