| Model Number |
20 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that when attempting to power on their device it would not complete a boot cycle.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that when attempting to power on their device it would not complete a boot cycle.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device but was unable to duplicate or verify the reported issue.As a precaution the power module and main capacitor were replaced.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|