Model Number 3660
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2020
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number 1627487-2020-31197.1627487-2020-31196.It was reported during the elective ipg replacement the leads extension were inserted in the header and in post -op the impedance was high so patient was taken back into the or were it was discovered that the one of the extension pull out of the ipg header the extension was reinserted resolving the issue.
 
Manufacturer Narrative
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key10515430
MDR Text Key206350424
Report Number3006705815-2020-31301
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public05415067031419
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/19/2022
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000095034
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/24/2020
Initial Date FDA Received09/10/2020
Supplement Dates Manufacturer Received11/12/2020
Supplement Dates FDA Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SCS EXTENSION; SCS EXTENSION
Patient Outcome(s) Other;
Patient Weight76
-
-