| Model Number |
2353-4105 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "t2 alpha tibial bone, disassembling device on distal device was not stable due to grazing, and it was confirmed that one pin was missing.The procedure was changed to radio root.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: refer to d9/h3.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿the stryker "instructions for cleaning, sterilization, inspection and maintenance" help to determine whether an instrument is in good condition or must be replaced due to excessive wear or obvious defects.Before the surgical procedure, ensure that all components prepared for the procedure function correctly with each other.During the procedure, repeatedly check that the connections between the instruments or between implants and instruments required for the precise positioning and fixing are secure.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.H3 other text : device disposition is unknown.
|
| |
|
Event or Problem Description
|
|
As reported: "t2 alpha tibial bone, disassembling device on distal device was not stable due to grazing, and it was confirmed that one pin was missing.The procedure was changed to radio root.".
|
| |
|