• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RATCHET WRENCH; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RATCHET WRENCH; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 71070367
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/19/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the ratchet wrench froze up and was no longer works.The issue happened outside the patient.The procedure was completed without delay, using a backup from smith & nephew to finish the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the device shows signs of extensive use.A functional evaluation confirmed the stated failure mode.The ratchet wrench has seized rendering the device inoperable.The device was manufactured in 2010.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RATCHET WRENCH
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10517618
Report Number1020279-2020-04519
Device Sequence Number3656737
Product Code KTT
UDI-Device Identifier00885556119174
UDI-Public00885556119174
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090347
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71070367
Device Catalogue Number71070367
Device Lot Number10LPM0012
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/19/2020
Supplement Date Received by Manufacturer12/15/2020
Initial Report FDA Received Date09/10/2020
Supplement Report FDA Received Date12/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-