• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000254
Device Problems Material Puncture/Hole (1504); Insufficient Flow or Under Infusion (2182); Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the staff received 3 possible helium loss 2 alarms in 10 minutes when the intra-aortic balloon (iab) was in use.No blood was noted in the driveline tubing, no visible kinks, and occasional ectopy.Patient's pressures have remained stable during all alarms and pauses in pumping.The staff received a high baseline alarm.Pumping resumed without issue.As a result, the iab was swapped out and the same insertion site was used.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
(b)(4).Teleflex received the device for investigation.The reported complaint of iab leak suspected is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.The returned device passed visual and functional test specifications.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
 
Event Description
It was reported that the staff received 3 possible helium loss 2 alarms in 10 minutes when the intra-aortic balloon (iab) was in use.No blood was noted in the driveline tubing, no visible kinks, and occasional ectopy.Patient's pressures have remained stable during all alarms and pauses in pumping.The staff received a high baseline alarm.Pumping resumed without issue.As a result, the iab was swapped out and the same insertion site was used.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10517675
MDR Text Key206496652
Report Number3010532612-2020-00248
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902007247
UDI-Public00801902007247
Combination Product (y/n)N
PMA/PMN Number
K021462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberIPN000254
Device Catalogue NumberIAB-05840-LWS
Device Lot Number18F20C0034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2020
Initial Date Manufacturer Received 08/21/2020
Initial Date FDA Received09/10/2020
Supplement Dates Manufacturer Received10/09/2020
Supplement Dates FDA Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-