MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAB
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump passed the displacement test but motor error alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime, compromised forced sensor system and excessive no delivery test due to motor error alarm.Motor passed motor test.(b)(4).
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had motor error.Customer reported that the drive support cap was normal.Customer was able to successfully clear the alarm and able to complete insulin pump rewind.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
MDR Report Key10519472
Report Number2032227-2020-175368
Device Sequence Number11488750
Product Code OZO
UDI-Device Identifier00643169503670
UDI-Public(01)00643169503670
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAB
Device Catalogue NumberMMT-751NAB
Device Lot NumberA4751NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2020
Initial Report FDA Received Date09/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age20 YR
Patient Weight80
-
-