MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735670
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Other relevant device(s) are: product id: 8801075, serial/lot #: (b)(4), ubd: 24-mar-2022, udi#: (b)(4).No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system being used in a procedure.It was reported that the 5 navigation spheres that were applied to the neuro navigation pointer were not being detected by the navigation arm.The reference array spheres were all being detected from the same lot number packages.The finish on the spheres looked as though it had been rubbed off or was much more matte than the other spheres.An instrument was used to pop them off of the navigation pointer resulting in some cut marks on the spheres but there were no abrasive in contact with the spheres before use, only surgical gloves during application.There was no person affected.The device is available for investigation.This was detected before use.
 
Additional Manufacturer Narrative
H3) the spheres used alongside the navigation system were returned to the manufacturer for evaluation.Testing found that there were cut marks on the units, noted to be inflicted during removal.Additionally, the spheres were noted to have been exposed to an abrasive that had removed an area of the reflective surface of each sphere.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 SYSTEM
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key10520320
Report Number1723170-2020-02416
Device Sequence Number3222323
Product Code HAW
UDI-Device Identifier00643169722194
UDI-Public00643169722194
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K162309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735670
Device Catalogue Number9735670
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer10/08/2020
Initial Report FDA Received Date09/11/2020
Supplement Report FDA Received Date10/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-