Section d4: multiple attempts were made to obtain additional information from the customer regarding the event (including device serial number); however, no additional information was provided.Manufacturer investigation conclusion: the reported event of the system controller being unable to communicate with the system monitor could not be confirmed as the system controller was not returned for analysis.Additional information provided stated that the issue happened with multiple system monitors.It was resolved by exchanging the system controller.Multiple attempts were made to obtain additional information from the customer regarding the event; however, no additional information was provided.The root cause of the reported event could not be conclusively determined through this analysis.Heartmate ii instructions for use section 4 ¿ ¿system monitor¿ explains how to use the system monitor with the system controller, and the actions to take when communication cannot be established.The patient handbook also cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|