• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Model Number 11451100
Device Problem Structural Problem (2506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Follow-up #1: date of submission 07/18/2017.Device evaluation: the pump has been returned and evaluated by product analysis on 06/20/2017 with the following findings: during investigation, the battery compartment was cracked above, and below the bumper pad.This report is made under the requirements of the medical device reporting regulations and does not constitute an admission on the part of animas of any deficiency in the performance of the device.
 
Event Description
On (b)(6) 2017, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was alleged that the battery compartment was damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
jon mulberg
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key10523319
MDR Text Key209616871
Report Number2531779-2020-00236
Device Sequence Number1
Product Code OYC
UDI-Device Identifier10840406100051
UDI-Public10840406100051
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number11451100
Device Catalogue Number11451100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2017
Was the Report Sent to FDA? Yes
Device Age21 MO
Date Manufacturer Received05/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
-
-