| Model Number |
PK-EFC0220T |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Multiple mdrs reported for the same patient event, see 3013680140-00002 & 3013680140-00003.During investigation, all manufacturing records were found to be conforming.Root cause identified to be engineering specification error which allowed manufacturing variability and ultimately a thin wall condition.This is the company's first complaint; we have been working with the esig and emdr teams to gain access and have a successful test submission over the past 2 months hence the delay in 30-day reporting.
|
| |
|
Event or Problem Description
|
|
It was reported that screws pulled through 3 of the epifix plates during surgery.It was reported that this did not affect the outcome of the surgery on the patient.No injury or delay of surgery was reported.3 plates were explanted, remaining 4 plates were implanted successfully.No allegations were reported against the screws or instruments in the procedural kits.This is device 1 of 3 (see 3013680140-2020-00002 & 3013680140-2020-00003 for remaining devices).Device was used for treatment, not diagnosis.
|
| |
|