| Model Number |
11451100 |
| Medical Device Problem Code |
Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission 03/23/2016 device evaluation: the device has been returned and evaluated by product analysis on 03/15/2016 with the following findings: a review of the pump¿s black box revealed a cs 052 alarm.The ez-prime steps were performed correctly with no alarms occurring during investigation.The product performed within specification.The original complaint was unable to be duplicated.The pump¿s cover was removed and there was no intermittent condition found to the printed circuit board or to the continuous glucose monitor module.Unrelated to the original complaint, the battery compartment was observed to be cracked.This report is made under the requirements of the medical device reporting regulations and does not constitute an admission on the part of animas of any deficiency in the performance of the device.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, the reporter contacted animas, alleging a call service alarm (cs 052/053/054/055 sleep) issue.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
| |
|