| Model Number |
45010 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a loss of aspiration.A 2.1mm jetstream xc catheter and jetstream console were selected for an atherectomy procedure in the left leg tibial popliteal trunk over a non-bsc filter.The motor drive unit sounded like it was making noise, however, aspiration was lost.The motor drive quit working.When the catheter was removed from the body, the exterior was observed to be kinked and aspiration stopped.The procedure was completed with another jetstream catheter.There were no patient complications reported and the patient did well.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a loss of aspiration.A 2.1mm jetstream xc catheter and jetstream console were selected for an atherectomy procedure in the left leg tibial popliteal trunk over a non-bsc filter.The motor drive unit sounded like it was making noise, however, aspiration was lost.The motor drive quit working.When the catheter was removed from the body, the exterior was observed to be kinked and aspiration stopped.The procedure was completed with another jetstream catheter.There were no patient complications reported and the patient did well.It was further reported that the console is being used with no issues.
|
| |
|