| Model Number |
68LP-400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/18/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿left side deflation and capsular contracture, baker grade ii¿.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: opening curved on anterior, creases fold and wear abrasion.Leak test and microscopic analysis was performed which identified: striated opening on anterior, opening crease sharp on posterior and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening on anterior assessed as surgical damage consist in the use of some surgical tool and a sharp creased opening on posterior assessed as fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|