| Model Number |
170052 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Dull, Blunt (2407); Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
08/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic sleeve gastrectomy, when the device was being applied at imbercation of the staple line after staple firing <(>&<)> resection of the greater curvature of the stomach, from the first pass of the needle reloads at the top of the stomach, an audible crunch could be heard.After few passes, the sutures became hung up within the tissue.The surgeon twisted, squeezed and pulled the device to free it from the tissue and as the needle finally passed through, there was additional bleeding from the stomach tissue.The surgeon requested for surgical snow to apply to the bleeding.The needles were blunt and there were also difficulty passing the needle into the receiving socket/hole.Tissue damage was noted.The surgeon continued to use the device to complete imbercating the rest of the staple line.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture was too tight, the device was difficult to toggle, and the tip of the needle was blunt or dull.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|