• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. JELCO SAFETY PROTECTIV PLUS CATHETER; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. JELCO SAFETY PROTECTIV PLUS CATHETER; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM L Back to Search Results
Model Number 306601
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
One jelco safety protective plus catheter was returned for analysis without the severed tube segment, needle guard assembly, sheath, or blister packaging.The sample was examined under magnification and photographed.The eyelet to tube securement area within the catheter hub was intact with no evidence of catheter securement issues.Examination of the catheter assembly noted that the catheter tube was severed at 1/16" above the catheter hub nose.Examination of the catheter hub displayed evidence of damage to the catheter hub nose.The manufacturing process was reviewed; an issue with low puck count in the pcm2 bevel module on a shift in may with no cause established.An attempt to replicate the damage found in the faulted sample was performed to another sample.It too was examined under magnification and photographed finding similarity to the complaint sample.The replicate sample also displayed evidence of a puncture in the catheter tubing, but the shape of the damage to the catheter tube differs from the complaint sample.Based on the evidence, the complaint was confirmed.The root cause was found due to manufacturing.
 
Event or Problem Description
Information was received indicating that upon removal of a smiths medical jelco safety protective plus catheter, the catheter broke off at the hub, leaving the tip in the cephalic vein.Vascular surgery was reported to remove the fragment in the operating room.There were no further reported adverse effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO SAFETY PROTECTIV PLUS CATHETER
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM L
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
MDR Report Key10532245
Report Number3012307300-2020-09282
Device Sequence Number11531409
Product Code FOZ
UDI-Device Identifier10351688071170
UDI-Public10351688071170
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K990236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date05/05/2023
Device Model Number306601
Device Catalogue Number306601
Device Lot Number3981802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2020
Initial Report FDA Received Date09/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient Weight125
-
-