CROSSTEX INTERNATIONAL INC. COMFORT PLUS PREMIUM SALIVA EJECTORS Back to Search Results
Model Number ZCB
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The user facility reported a saliva ejector tip detached from the comfort plus premium saliva ejector tube during a dental procedure.The tip was not swallowed by the patient.The patient was under local anesthetic during the procedure when the saliva ejector tip was detached.It was reported the dentist performing the procedure was able to retrieve the saliva ejector tip from the patients mouth before it was swallowed.The reported event did not cause or contribute any serious injury or deterioration of health.This event could potentially lead to serious injury if it were to recur.This complaint will be monitored in the crosstex complaint handling system.
 
Event or Problem Description
The user facility reported a saliva ejector tip detached from the comfort plus premium saliva ejector tube during a dental procedure.The tip was not swallowed by the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT PLUS PREMIUM SALIVA EJECTORS
Common Device Name
SALIVA EJECTORS
Manufacturer (Section D)
CROSSTEX INTERNATIONAL INC.
10 ranick road
happauge NY 11788
MDR Report Key10532477
Report Number2433773-2020-00004
Device Sequence Number14216825
Product Code DYN
UDI-Device Identifier50732224000676
UDI-Public(01)50732224000676
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZCB
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2020
Initial Report FDA Received Date09/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-