H10: b4: event description updated h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.The arthroscopic photo provided confirmed the reported failure.The complaint was confirmed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the process summary, form, fill and seal packaging sequence for: arthroscopic surgical blades products, found that the blade device is to be inspected for component integrity and particulate or foreign matter.A review of the complaint revealed it was not confirmed if all the metal debris was completely removed.Therefore, since the metal debris from the elite disposable blade is not implantable, we cannot rule out the possibility of tissue inflammation, micro-motion or migration of the metal debris.Per report the procedure completed with the same device, less than a thirty-minute delay and there was no other harm to the patient.Therefore, no further clinical/medical assessment is warranted at this time.No containment or corrective actions are recommended at this time.
|