• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND SENTRANT INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND SENTRANT INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number SENSH1428W
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A sentrant sheath was intended to be used as an accessory device in an unknown endovascular procedure.It was reported that prior to the procedure, the sheath hub and valves pulled apart while the device was being prepped for the case.A replacement sheath was used to in the procedure instead.No cause of the event was reported.No additional clinical sequelae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Product analysis conclusion;photo received from the account confirmed the seal cap detachment as reported.The sentrant sheath returned with the dilator loaded and the seal cap and seal detached from the seal housing.The seal cap and seal were removed on the dilator.Inspection of the seal cap tabs confirmed an indentation on both tabs.The tabs appeared to have been orientated incorrectly in the window.The reported detachment was confirmed through analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTRANT INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
MDR Report Key10533540
Report Number9612164-2020-03455
Device Sequence Number13546324
Product Code DYB
UDI-Device Identifier00643169792586
UDI-Public00643169792586
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K123990
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2022
Device Model NumberSENSH1428W
Device Catalogue NumberSENSH1428W
Device Lot Number00151114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2020
Initial Date Received by Manufacturer 08/19/2020
Supplement Date Received by Manufacturer11/10/2020
Initial Report FDA Received Date09/15/2020
Supplement Report FDA Received Date11/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-