• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS SHENZHEN MAGNETIC RESONANCE LTD. ARTIS ONE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS SHENZHEN MAGNETIC RESONANCE LTD. ARTIS ONE; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10848600
Medical Device Problem Codes Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that a malfunction occurred while operating the artis one system.During a cardiology procedure, the user reported that the monitor display in the exam room started to blink (image appearing and disappearing).The user restarted the system ,but couldn't fix the problem.The patient was transferred to another system in the same hospital, and finished the procedure successfully.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation was performed considering complaint description, cs reports, system history and system log files.The investigation showed that the cause of the error was due to a graphic card defect.According to information from field, during s cardiology procedure the monitor display in the exam room started to blink (image appearing and disappearing).The customer attempted to restart the system but couldn't fix the problem.The customer service engineer (cse) on site checked and replaced the graphic card of the smart pc.After the hardware exchange, the system recovered and works as specified.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ONE
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.
gaoxin c ave.
2nd, hi-tech industrial park
shenzhen, china 51805 7
CH  518057
MDR Report Key10534046
Report Number3004754211-2020-44310
Device Sequence Number3735395
Product Code OWB
UDI-Device Identifier04056869009964
UDI-Public04056869009964
Combination Product (Y/N)N
Initial Reporter CountryMO
PMA/510(K) Number
K133580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10848600
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/08/2020
Supplement Date Received by Manufacturer10/13/2020
Initial Report FDA Received Date09/15/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-