| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device bla number: na.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -14.00 diopter, in the patients right eye (od) on (b)(6) 2020.The lens was explanted on (b)(6) 2020 due to high pressure.The lens was exchanged for a different size lens.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3: device evaluation: the lens was returned dry, in a specimen cup, with residue on lens.Visual inspection found no visible damage to the lens and there was residue on lens.H6: health impact- clinical code: 4581 - angle closure.H6: health effect impact code: 4625 - additional surgery; yag - addition of new pi.Anterior chamber irrigation/evacuation of visco/fluids.Corrected data: b5: the reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -14.00 diopter, in the patient's left eye (os) on (b)(6) 2020.The lens was explanted on (b)(6) 2020 due to excessive vaulting, angle closure with elevated intraocular pressure.Additional surgery was performed, addition of new pi by yag and anterior chamber irrigation/evacuation of visco/fluids.The lens was exchanged for a shorter lens and the problem was not resolved.See mfr report # 2023826-2020-02119 for exchanged lens.The cause of the event was due to patient-related factor.Claim # (b)(4).
|
| |
|