| Model Number |
71430401 |
| Medical Device Problem Codes |
Break (1069); Flaked (1246); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that in the sterilization process department was noted a small piece of plastic spacer insert chipped off.As a result, the locking mechanism of the trial insert has been damaged and can no longer be used.There are not any patient involved in the case.The procedure performed before the sterilization process was completed without delay and with a backup from smith nephew.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated lot number and exp date and udi and mnf date.
|
| |
|
Event or Problem Description
|
|
It was reported that in the sterilization process department was noted a small piece of plastic spacer insert chipped off.As a result, the locking mechanism of the trial insert has been damaged and can no longer be used.There aren´t any patient involved in the case.The procedure performed before the sterilization process was completed without delay and with a backup from smith nephew.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be chipped and causing the locking mechanism to seized.The device shows signs of extensive use.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|