MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM; PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER Back to Search Results
Model Number 74027214
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/11/2013
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the clinical subject suffered from failure to progress with flexion.The event was treated via manipulation under anesthesia and was resolved.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this clinical study adverse event form reports a manipulation under anesthesia due to right knee stiffness.The requested x-rays and surgical reports have not been provided to date.Therefore, based on limited information, no further assessment can be performed at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated at that time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes could be alignment, fit/size of device used or wear.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM
Common Device Name
PRSTHSS,KNEE,PTLLFMRTBIAL,SM-CNSTRAIND,CMNTD,POLYMER/MTL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10538525
Report Number1020279-2020-04595
Device Sequence Number3872897
Product Code JWH
UDI-Device Identifier00885556178812
UDI-Public00885556178812
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/14/2021
Device Model Number74027214
Device Catalogue Number74027214
Device Lot Number11LM07511
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/17/2020
Supplement Date Received by Manufacturer06/29/2021
Initial Report FDA Received Date09/15/2020
Supplement Report FDA Received Date06/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-