• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAL
Medical Device Problem Codes No Display/Image (1183); Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pump received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify unexpected battery power loss anomaly due to blank display.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump screen went blank, they were able to get it back and received battery out limit and bolus stopped alarm.They were able to clear the alarm.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
MDR Report Key10539392
Report Number2032227-2020-177725
Device Sequence Number571331
Product Code OYC
UDI-Device Identifier00643169507180
UDI-Public(01)00643169507180
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAL
Device Catalogue NumberMMT-523NAL
Device Lot NumberB4523NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2020
Initial Report FDA Received Date09/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age42 YR
Patient Weight70
-
-