| Model Number |
3602536 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis#: 703805582:3602536, lot# rm18a013.Visual and optical examination identified that the hex of the cone set screw has been stripped.The stripped cone set screw is consistent with over-torque.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider via a manufacturer representative regarding a patient with trauma.It was reported that the lateral set screw stripped as surgeon was attempting to final tighten.The device was replaced, and will be returned.There were no symptoms reported.There were no further complications reported regarding the event.On (b)(6) 2020, (rep) additional information was received from manufacturer representative regarding a patient with occiput to c4 p osterior fusion (levels implanted-occiput to c4) due to trauma.
|
| |
|